M14 Firmware Versions ***************************** 00x.001 - first version 00x.002 - fast relay response at peak with 0 s delay time 00x.003 - bug in power-on delay at initial overload fixed - OVL LED flashes also at RMS/peak polling 00x.004 - M14-WEB issue of version 003 fixed - occasional overloads by integrator overflow fixed - false alarms with 11.5 kHz low pass fixed 00x.005 - lower noise with high pass 0.3/5/10 Hz for acceleration 00x.006 - support for new hardware M14B - four-state IEPE LED - improved noise performance with high pass 0.3 Hz - bar graph color at teach in 00x.007 - RS-485 interface supports MODBUS-RTU - new USB command #Q for baudrate - bug in command #N fixed 00x.008 - MODBUS bug in peak values fixed - MODBUS bug in filter settings fixed 00x.009 - normal operation of 4-20 mA with switched off IEPE 00x.010 - bug fixed in peak measurement 00x.011 - MODBUS command set and error reporting changed - baud rate and MODBUS address added to USB command #X